THIOCYANATE ION-SELECTIVE PVC MEMBRANE ELECTRODE
Date
2010
Journal Title
Journal ISSN
Volume Title
Publisher
Academia de Știinițe a Moldovei
Abstract
A potentiometric selective sensor based on trinuclear chromium(III) complex as a novel ionophore for the thiocyanate-selective electrode is reported. The sensor displays a near Nernstian slope of 57 ± 2 mV per decade, over a wide pH range 3 - 11. The working concentration range of the electrode is 1.10-5 – 1.10-1 mol/L with a detection limit of 5.
10-6 mol/L. The sensor has a response time of 20 s and can be used for at least 6 months without any considerable fluctuation of the potential. The selectivity coeffi cients determined at using the fi xed interference method indicate a good discriminating ability towards other anions. The prepared sensor was applied as an indicator electrode in the
titration of thiocyanate with Ag+.
Description
Keywords
thiocyanate, potentiometric sensor, titration
Citation
REVENCO, M. , MARTIN. M. et al. Thiocyanate ion-selective pvc membrane electrode. In: Cemistry Journal of Moldova. 2010, nr.1, pp. 73-77. ISSN1857-1727.