THIOCYANATE ION-SELECTIVE PVC MEMBRANE ELECTRODE
dc.contributor.author | Revenco, Mihail | |
dc.contributor.author | Martin, Mariana | |
dc.contributor.author | Dayyih, Waell Abu | |
dc.date.accessioned | 2018-03-27T12:06:20Z | |
dc.date.available | 2018-03-27T12:06:20Z | |
dc.date.issued | 2010 | |
dc.description.abstract | A potentiometric selective sensor based on trinuclear chromium(III) complex as a novel ionophore for the thiocyanate-selective electrode is reported. The sensor displays a near Nernstian slope of 57 ± 2 mV per decade, over a wide pH range 3 - 11. The working concentration range of the electrode is 1.10-5 – 1.10-1 mol/L with a detection limit of 5. 10-6 mol/L. The sensor has a response time of 20 s and can be used for at least 6 months without any considerable fluctuation of the potential. The selectivity coeffi cients determined at using the fi xed interference method indicate a good discriminating ability towards other anions. The prepared sensor was applied as an indicator electrode in the titration of thiocyanate with Ag+. | en |
dc.identifier.citation | REVENCO, M. , MARTIN. M. et al. Thiocyanate ion-selective pvc membrane electrode. In: Cemistry Journal of Moldova. 2010, nr.1, pp. 73-77. ISSN1857-1727. | en |
dc.identifier.issn | 1857-1727 | |
dc.identifier.uri | https://msuir.usm.md/handle/123456789/1724 | |
dc.language.iso | en | en |
dc.publisher | Academia de Știinițe a Moldovei | en |
dc.subject | thiocyanate | en |
dc.subject | potentiometric sensor | en |
dc.subject | titration | en |
dc.title | THIOCYANATE ION-SELECTIVE PVC MEMBRANE ELECTRODE | en |
dc.type | Article | en |
Files
Original bundle
1 - 1 of 1
- Name:
- Thiocyanate ion-selective pvc membrane electrode73-77.pdf
- Size:
- 105.28 KB
- Format:
- Adobe Portable Document Format
- Description:
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.71 KB
- Format:
- Item-specific license agreed upon to submission
- Description: