Thiocyanate ion-selective PVC membrane electrode [Articol]
dc.contributor.author | Revenco, Mihail | |
dc.contributor.author | Martin, Mariana | |
dc.contributor.author | Dayyih, Waell Abu | |
dc.date.accessioned | 2016-10-26T15:44:33Z | |
dc.date.available | 2016-10-26T15:44:33Z | |
dc.date.issued | 2010 | |
dc.description.abstract | A potentiometric selective sensor based on trinuclear chromium(III) complex as a novel ionophore for the thiocyanate-selective electrode is reported. The sensor displays a near Nernstian slope of 57 ± 2 mV per decade, over a wide pH range 3 - 11. The working concentration range of the electrode is 1.10-5 – 1.10-1 mol/L with a detection limit of 5.10-6 mol/L. The sensor has a response time of 20 s and can be used for at least 6 months without any considerable fluctuation of the potential. The selectivity coefficients determined at using the fixed interference method indicate a good discriminating ability towards other anions. The prepared sensor was applied as an indicator electrode in the titration of thiocyanate with Ag+. | en |
dc.identifier.citation | REVENCO, Mihail, MARTIN, Mariana, DAYYIH, Waell Abu. Thiocyanate ion-selective PVC membrane electrode. In: Chemistry Journal of Moldova: General, Industrial and Ecological Chemistry. 2010, nr. 5 (1), pp. 73-77. | en |
dc.identifier.issn | 1857-1727 | |
dc.identifier.uri | https://doaj.org/toc/2345-1688 | |
dc.identifier.uri | https://msuir.usm.md/handle/123456789/929 | |
dc.language.iso | en | en |
dc.publisher | Institute of Chemistry of ASM | en |
dc.subject | thiocyanate | en |
dc.subject | potentiometric sensor | en |
dc.subject | titration | en |
dc.title | Thiocyanate ion-selective PVC membrane electrode [Articol] | en |
dc.type | Article | en |