Roentgen image recording on sillenite-type electrooptical crystals by electrophotographic method [Articol]
Date
2005
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
To record Roentgen image on sillenite-type crystals while charging in a
corona discharge field, a device has been performed. The resolution not less than 10 pairs of lines per mm, the sensitivity of 25 reciprocal Roentgens, the contrast coefficient not less than 2 at the exposure length of 0.6 s were obtained using Bi12GeO20crystal. The device survives more than 104 of writing-erase
cycles. From the sensitivity point of view, the considered registration is just as good as X-Ray films used for radiography and dosimetry.
Description
Keywords
radiography
Citation
DEMENTIEV, Ig., MAXIMOV, E. et al. Roentgen image recording on sillenite-type electrooptical crystals by electrophotographic method. In: Moldavian Journal of the Physical Sciences. 2005, nr. 1, pp. 64-68. ISSN 1810-648X.